In vivo evaluation of focused ultrasound ablation surgery (FUAS)-induced coagulation using echo amplitudes of the therapeutic focused ultrasound transducer

ORCID Icon, &
Article: 2325477 | Received 15 Jan 2024, Accepted 26 Feb 2024, Published online: 04 Mar 2024
 

Abstract

Objective

Monitoring sensitivity of sonography in focused ultrasound ablation surgery (FUAS) is limited (no hyperechoes in ∼50% of successful coagulation in uterine fibroids). A more accurate and sensitive approach is required.

Method

The echo amplitudes of the focused ultrasound (FUS) transducer in a testing mode (short pulse duration and low power) were found to correlate with the ex vivo coagulation. To further evaluate its coagulation prediction capabilities, in vivo experiments were carried out. The liver, kidney, and leg muscles of three adult goats were treated using clinical FUAS settings, and the echo amplitude of the FUS transducer and grayscale in sonography before and after FUAS were collected. On day 7, animals were sacrificed humanely, and the treated tissues were dissected to expose the lesion. Echo amplitude changes and lesion areas were analyzed statistically, as were the coagulation prediction metrics.

Results

The echo amplitude changes of the FUS transducer correlate well with the lesion areas in the liver (R = 0.682). Its prediction in accuracy (94.4% vs. 50%), sensitivity (92.9% vs. 35.7%), and negative prediction (80% vs. 30.8%) is better than sonography, but similar in specificity (80% vs. 100%) and positive prediction (100% vs. 100%). In addition, the correlation between tissue depth and the lesion area is not good (|R| < 0.2). Prediction performances in kidney and leg muscles are similar.

Conclusion

The FUS echo amplitudes are sensitive to the tissue properties and their changes after FUAS. They are sensitive and reliable in evaluating and predicting FUAS outcomes.

Acknowledgments

The authors thank Dr. Tianfeng Zhang for experimental work.

Ethical approval

This study was approved by the Institutional Animal Care and Use Committee (IACUC) of Chongqing Haifu Hospital (approval no. 2022-011, Chongqing, China).

Disclosure statement

No potential conflict of interest was reported by the author(s).

Data availability statement

The data presented in this study are available on request from the corresponding author.

Additional information

Funding

This work was supported by the Future Innovative Research Fund of Chongqing Medical University (2022-W0063).