Preclinical Investigation of Ab Interno Goniotomy Using Three Different Techniques

ORCID Icon, &
Pages 2619-2623 | Received 08 Jun 2023, Accepted 17 Aug 2023, Published online: 04 Sep 2023
 

Abstract

Purpose

To evaluate incisional or excisional tissue-level effects of ab interno goniotomy techniques on human trabecular meshwork (TM).

Methods

The TM from human cadaveric corneal rim tissue was treated using three devices: (1) Kahook Dual Blade (KDB) GLIDE, (2) iAccess, and (3) SION. Two human corneal rims were used for each of the iAccess and SION devices and one with the KDB GLIDE, with 360 degrees of TM treated in each case. Sections were then prepared for analysis and comparison between devices. Tissue samples underwent standard histologic processing with H&E stain, followed by comparative analyses.

Results

Areas treated with the KDB GLIDE device resulted in nearly complete excision of TM overlying the canal of Schlemm without injury to surrounding tissues. The iAccess device can be used as a focal trephine to create holes or dragged for TM disruption. When used to create holes, iAccess punched through the full thickness of the TM and also disrupted the anterior scleral tissue. It caused some incisional openings through the TM but with significant leaflets remaining and minimal true “hole-punch” effect. When the device tip was dragged, iAccess incised the TM and left debris behind with little, if any, excision of tissue. SION led to both incision and excision of TM with incision predominating over excision.

Conclusion

The various methods evaluated to perform ab interno goniotomy resulted in varying degrees of TM incision or excision. Only the KDB GLIDE device resulted in reliable excision of TM, while the other devices produced incision or minimal excision of tissue with residual leaflets and debris. Use of iAccess resulted in focal disruption of the anterior scleral wall. Because incisional approaches that leave longer residual leaflets may be more prone to fibrosis and closure compared to excisional treatments, clinical correlation will be necessary to better understand the significance of these findings with respect to relative effectiveness of intraocular pressure lowering in eyes with glaucoma.

Disclosure

Dr David A Ammar reports Contract Research from Lions World Vision Institute, during the conduct of the study; Contract Research from Lions World Vision Institute, outside of the submitted work. Mr Eric Porteous is an employee of New World Medical. Dr Malik Kahook reports being a consultant to New World Medical during the conduct of the study. In addition, Dr Malik Kahook has a patent US10,327,947B2 issued to and owned by New World Medical.