136
Views
0
CrossRef citations to date
0
Altmetric
CLINICAL TRIAL REPORT

The Significance of Dry Eye Signs on Preoperative Keratometry Measurements in Patients Scheduled for Cataract Surgery

ORCID Icon, ORCID Icon, ORCID Icon, ORCID Icon, ORCID Icon, ORCID Icon & ORCID Icon show all
Pages 151-161 | Received 17 Nov 2023, Accepted 08 Jan 2024, Published online: 16 Jan 2024
 

Abstract

Purpose

The primary objective was to investigate if subjects with dry eyes had increased variability of keratometry measurements prior to cataract surgery compared to subjects with non-dry eyes. Secondary objectives were to determine which separate signs affected keratometry.

Patients and Methods

This study was part of a prospective interventional randomized controlled trial. After dry eye diagnostics were performed (signs only) subjects were divided into sign of dry eye (SDE) positive and negative groups. To investigate variability, we performed two keratometry measurements for each subject with three different optical biometers: Anterion (OCT optical biometer), Eyestar (combined OCT and reflection-based optical biometer), and Lenstar (reflection based-optical biometer).

Results

One hundred and thirty-one subjects were available for analysis. The variability of astigmatism was significantly higher for subjects with hyperosmolarity compared to normal eyes for the Lenstar, as was the percentage of eyes with variability of astigmatism greater than 0.25 D. The percentage of eyes with variability of average K greater than 0.25 D was higher for subjects with non-invasive keratograph break-up time <10 seconds (NIKBUT positive) compared to normal eyes for the Lenstar.

Conclusion

Combined diagnostic criteria (signs only) showed no statistically significant differences for keratometry measurements between SDE positive and negative. Eyes with hyperosmolarity and NIKBUT positive showed statistically higher variability of keratometry measurements compared to normal eyes for Lenstar, but not for the Anterion or Eyestar biometers.

Disclosure

Richard (Rick) Potvin reports consulting fees from Alcon and Hoya. The other authors report no conflict of interest in this article.