1,470
Views
0
CrossRef citations to date
0
Altmetric
RESEARCH ARTICLE

Overset meshing in combination with novel blended weak-strong fluid-structure interactions for simulations of a translating valve in series with a second valve

, , , , &
Pages 736-750 | Received 01 Jun 2022, Accepted 30 Mar 2023, Published online: 18 Apr 2023
 

Abstract

Mechanical circulatory support (MCS) devices can bridge the gap to transplant whilst awaiting a viable donor heart. The Realheart Total Artificial Heart is a novel positive-displacement MCS that generates pulsatile flow via bileaflet mechanical valves. This study developed a combined computational fluid dynamics and fluid-structure interaction (FSI) methodology for simulating positive displacement bileaflet valves. Overset meshing discretised the fluid domain, and a blended weak-strong coupling FSI algorithm was combined with variable time-stepping. Four operating conditions of relevant stroke lengths and rates were assessed. The results demonstrated this modelling strategy is stable and efficient for modelling positive-displacement artificial hearts.

Disclosure statement

AN and ILP are employees of, or consultants to, and/or shareholders of, Scandinavian Real Heart AB.

Additional information

Funding

JB PhD funded 50/50 from EPSRC and Scandinavian Real Heart AB (Reference: 2426107)