200
Views
0
CrossRef citations to date
0
Altmetric
Device Profile

A profile on the WISE cortical strip for intraoperative neurophysiological monitoring

ORCID Icon &
Pages 373-379 | Received 20 Nov 2023, Accepted 11 Apr 2024, Published online: 17 Apr 2024

Figures & data

Table 1. Devices that are commercially available.

Figure 1. ECoG electrodes in the surgical field.

The WCS electrode (right) has dark platinum electrode contacts embedded in the carrier material. A standard electrode (Ad-Tech, left) has electrode contacts that are platelets made from platinum foil [Citation30].
Figure 1. ECoG electrodes in the surgical field.

Figure 2. Impedance and noise level. (a) The median impedance was 6.9 kΩ for Ad-Tech and 3.4 kΩ for WCS measured at 140 Hz.

(a) Linear spectral density (LSD) of the raw signal. Across all frequencies, the LSD of the WCS electrode (blue) was higher than the LSD of the Ad-Tech electrode (red).
(b) Example of an evoked HFO event with small amplitude. The small amplitude did not exceed the noise level of the Ad-Tech (blue lines), resulting in (SNR = 0.9 < 1) and the event went undetected. With the reduced noise level of WISE (red lines), the SNR increased to SNR = 1.6 > 1 and the HFO was detected. Adapted from [Citation30].
Figure 2. Impedance and noise level. (a) The median impedance was 6.9 kΩ for Ad-Tech and 3.4 kΩ for WCS measured at 140 Hz.